www.milwaukeeinst.com1www.milwaukeeinst.comMA871INSTRUCTION MANUALMilwaukee RefractometerMilwaukee RefractometerMilwaukee RefractometerMilwaukee Refra
1010101010Instruction Manual MA871 Refractometer for Sucrose MeasurementsMEASUREMENT PROCEDUREVerify the instrument has been calibrated before taking
www.milwaukeeinst.com115. Using plastic pipettes, rinse prism and sample well with distilled or deionized water.Wipe dry. The instrument is ready for
1212121212Instruction Manual MA871 Refractometer for Sucrose MeasurementsTo change the temperature measurement unit from Celsius to Fahrenheit (or vic
www.milwaukeeinst.com13ERROR MESSAGESedoCrorrEnoitpircseDrrEelcyC.eruliaflareneGfI.tnemurtsniotrewop,rorresahllitstnemurtsni.eekuawliMtcatnocOL yalpsi
1414141414Instruction Manual MA871 Refractometer for Sucrose MeasurementsBATTERY REPLACEMENTTo replace the instrument’s battery, follow these steps:•
www.milwaukeeinst.com15WARRANTYThis instrument is warranted against defects in materials and manufacturing for a period of2 years from the date of pur
1616161616Instruction Manual MA871 Refractometer for Sucrose MeasurementsSales and Technical Service Contacts:Milwaukee s.r.l.C.so Leonardo da Vinci 4
22222Instruction Manual MA871 Refractometer for Sucrose MeasurementsTABLE OF CONTENTSFUNCTIONAL DESCRIPTION ...
www.milwaukeeinst.com3FUNCTIONAL DESCRIPTIONDISPLAYDISPLAYDISPLAYDISPLAYDISPLAYA. BATTERY STATUS ICON (BLINKS WHEN LOW BATTERY CONDITION DETECTED)B. M
44444Instruction Manual MA871 Refractometer for Sucrose MeasurementsH. BATTERY COVERI. BATTERY COMPARTMENTBOTTOMBOTTOMBOTTOMBOTTOMBOTTOMA. LIQUID CRYS
www.milwaukeeinst.com5GENERAL DESCRIPTIONSIGNIFICANCE OF USEThank you for choosing Milwaukee. This instruction manual will provide you the necessaryin
66666Instruction Manual MA871 Refractometer for Sucrose MeasurementsRange 0 to 85% Brix 0 to 80 °C (32 to 175 °F)Resolution 0.1% Brix 0.1 °C (0.1 °F)A
www.milwaukeeinst.com7PRINCIPLE OF OPERATIONThe Brix determination is made by measuring the refractive index of a solution. RefractiveIndex is an opti
88888Instruction Manual MA871 Refractometer for Sucrose MeasurementsMEASUREMENT GUIDELINES• Handle instrument carefully. Do not drop.• Do not immerse
www.milwaukeeinst.com9CALIBRATION PROCEDURECalibration should be performed daily, before measurements are made, when thebattery has been replaced, or